Felix Instruments Inc. - a subsidiary of CID Bio-Science, Inc.

FelixModel F-940 -Ripening Gas Analyzer

SHARE

The F-940 Store It! Gas Analyzer provides rapid and precise measurements of ethylene, CO2 and O2. The Store It! was designed for Quality Assurance staff to increase shelf-life and prevent spoilage by monitoring ripening gases throughout the storage process.  The Store It! features an intuitive user interface, making it simple to operate right out of the box. Its lightweight, handheld design enables users to rely on the Store It! at all stages of storage along the supply chain. A built-in data logger stores thousands of data points and conveniently communicates data through USB or Bluetooth onto a PC or tablet for later analysis.

Most popular related searches
  • Measures C2H4, CO2 and O2
  • Displays results in under 30 seconds
  • Past records stored on the device
  • Handheld and lightweight
  • Internal data logging and storage (4GB SD card)
  • Bluetooth data transfer
  • Battery lasts for 8+ hours
  • Display easily viewed in dimly lit environments
  • PolarCept!TM water filter for removing  non-ethylene hydrocarbons
  • Controlled Atmosphere storage facility management
  • Ethylene mitigation verification
  • Packing house inspection
  • Precision citrus degreening
  • Storage assignment in distribution centers
  • Packaging quality assurance
  • Sampling Flow Rate: 70mL/min
  • Total Sampling Time: 30 seconds
  • Data Saving: Automated, 1 second intervals
  • Display: Backlit visible transflective LCD
  • Operating Environment: 0° C - 50° C (0-90 % humidity non-condensing)
  • Operating Atmospheric Pressure: 500-1200mbar
  • Dimensions: 18 cm x 13.5 cm x 5.5 cm
  • Weight: 0.95 kg
  • Enclosure: Powder coated aluminum
  • Battery Life: 8.5 hours
  • Power Source: Removable rechargeable lithium-ion battery
  • PC Interface: USB and WI-FI SD card + Bluetooth
  • Data Recorded with Each Measurement: Ethylene, CO2 and O2 concentrations, date, time, RH, GPS location

Trigger Mode Measurements

  • Sensor Type: Electrochemical Infrared Sensor, Pyroelectric detector Electrochemical
  • Range: 0-10ppm 0-100% 0-100%
  • Resolution: 0.01ppm 0.01% 0.10%
  • Lower Detection Limit: 0.15ppm 0.50% 0.30%
  • Sampling Volume: 30 ml 12 ml 12 ml
  • Response Time: 30 seconds 12 seconds 12 seconds
  • Accuracy (absolute): 0.15 ppm 0.50% 0.30%
  • Accuracy (relative): 5% 3% 2%
  • Offset Calibration: Weekly Weekly Weekly
  • Span Calibration: 3 months 12 months 6 months
  • Sensor Replaceable (by user): yes no yes
  • Lifetime: 2 years 5 years 2 years

Continuous Mode Measurements
      

  • Sensor Type: Electrochemical Infrared Sensor, Pyroelectric detector Electrochemical
  • Range: 0-10 ppm 0-100% 0-100%
  • Resolution: 0.010 ppm 0.01% 0.10%
  • Lower Detection Limit: 0.030 ppm 0.01% 0.10%
  • Response Time: 70 seconds 6 seconds 12 seconds
  • Accuracy (absolute): 0.030 ppm 0.01% 0.10%
  • Accuracy (relative): 5% 3% 2%
  • Offset Calibration: Weekly Weekly Weekly
  • Span Calibration: 3 months 12 months 6 months
  • Sensor Replaceable (by user): yes no yes
  • Lifetime: 2 years 5 years 2 years